User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Doric 400A:Digital Temperature Indicator(17-63) DATE: 15-Aug-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 12 NUMBER OF LINES: 97 CONFIGURATION: Fluke 5700A STANDARD: Omega TRC III ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # This procedure is run with the MET/CAL Configuration #1 loaded. 1.001 ASK- R N P F W 1.002 ASK+ X 1.003 STD Omega TRC III 1.004 HEAD ADDITIONAL REQUIRED EQUIPMENT 1.005 DISP Omega TRC III Ice Point Cell 1.005 DISP Omega Type "T" Reference Thermocouple 1.006 HEAD INITIAL CONDITIONS 1.007 DISP Insert the reference thermocouple into the center 1.007 DISP well of the Ice Point Cell. 1.008 DISP Connect the copper end (white connector) to the 1.008 DISP 5700 OUTPUT HI/LO. Red wire is HI, green wire is LO. 1.009 DISP Connect the blue thermocouple side of the reference 1.009 DISP probe the the UUT. Use adaptors as necessary. 1.010 HEAD {* T/C Type T °C *} 1.011 JMP 2.001 1.012 EVAL 2.001 5700 -5.439mV S 2W 2.002 MEMI Use the keyboard to enter the UUT reading in °C. 2.003 ACC -190.0°C 0.1U 2.004 MEME 2.005 MEMC °C 1.0U #! Test Tol 1, Sys Tol 0.1, TUR 10.000 (>= 4.00). 3.001 5700 -3.379mV S 2W 3.002 MEMI Use the keyboard to enter the UUT reading in °C. 3.003 ACC -100.0°C 0.1U 3.004 MEME 3.005 MEMC °C 1.0U #! Test Tol 1, Sys Tol 0.1, TUR 10.000 (>= 4.00). 4.001 5700 0.000mV S 2W 4.002 MEMI Use the keyboard to enter the UUT reading in °C. 4.003 ACC 0.0°C 0.1U 4.004 MEME 4.005 MEMC °C 1.0U #! Test Tol 1, Sys Tol 0.1, TUR 10.000 (>= 4.00). 5.001 5700 4.279mV S 2W 5.002 MEMI Use the keyboard to enter the UUT reading in °C. 5.003 ACC 100.0°C 0.1U 5.004 MEME 5.005 MEMC °C 1.0U #! Test Tol 1, Sys Tol 0.1, TUR 10.000 (>= 4.00). 6.001 5700 6.704mV S 2W 6.002 MEMI Use the keyboard to enter the UUT reading in °C. 6.003 ACC 150.0°C 0.1U 6.004 MEME 6.005 MEMC °C 1.0U #! Test Tol 1, Sys Tol 0.1, TUR 10.000 (>= 4.00). 7.001 5700 9.288mV S 2W 7.002 MEMI Use the keyboard to enter the UUT reading in °C. 7.003 ACC 200°C 0.1U 7.004 MEME 7.005 MEMC °C 1.0U #! Test Tol 1, Sys Tol 0.1, TUR 10.000 (>= 4.00). 8.001 5700 12.013mV S 2W 8.002 MEMI Use the keyboard to enter the UUT reading in °C. 8.003 ACC 250.0°C 0.1U 8.004 MEME 8.005 MEMC °C 1.0U #! Test Tol 1, Sys Tol 0.1, TUR 10.000 (>= 4.00). 9.001 5700 14.862mV S 2W 9.002 MEMI Use the keyboard to enter the UUT reading in °C. 9.003 ACC 300.0°C 0.1U 9.004 MEME 9.005 MEMC °C 1.0U #! Test Tol 1, Sys Tol 0.1, TUR 10.000 (>= 4.00). 10.001 5700 17.819mV S 2W 10.002 MEMI Use the keyboard to enter the UUT reading in °C. 10.003 ACC 350.0°C 0.1U 10.004 MEME 10.005 MEMC °C 1.0U #! Test Tol 1, Sys Tol 0.1, TUR 10.000 (>= 4.00). 11.001 5700 20.255mV S 2W 11.002 MEMI Use the keyboard to enter the UUT reading in °C. 11.003 ACC 390.0°C 0.1U 11.004 MEME 11.005 MEMC °C 1.0U #! Test Tol 1, Sys Tol 0.1, TUR 10.000 (>= 4.00). 12.001 HEAD {} 12.002 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.